The 4012 Very Fast TLP (VFTLP+ ™) system is built with specially designed high speed hardware. Its test pulse simulates the CDM speed and its current and voltage sensors capture the very fast response of ESD circuits necessary for CDM design.
Test Pulse width: 1ns, 2ns, 5ns and 10ns (standard). Pulse width is software selectable.
Test Pulse voltage: 0-500v @ 50Ω load, 1,000v @ open circuit
Test Pulse current: 0-10 amps @ 50Ω load, 20 amps @ short circuit
Test Pulse risetime: 100ps, 200ps, 400ps (built-in, software selectable)
Leakage voltage: 0 to 100v (0.1 v increments)
Leakage current sensitivity: 1 pA to 2.5 mA